RCR-600 6-Point 100mm Chart Recorder

Standard Functions | Specifications | How to Order


Agency Approvals:

     
Design Features
  • 6-Channel dotting recorder
  • 100 mm chart paper size
  • 144 x 144 mm metal housing
  • Weighs only 3.3 lb. (1.5 Kg)
  • NEMA 4 / IP65 Dustproof water resistant housing
  • Universal settable input and range
  • Optional 6 alarm-relay outputs
  • Optional 3 digital inputs
  • Optional communication interface for RS-232

Standard Functions
Function Description Function Description
Analog Recording Makes analog recording with 6 colored dots. Self Diagnostics Indicates “Error” and code when there is a fault.
Digital Display Indicates channel number, process variable, date, chart speed and alarm setpoint. Open Input Indication Sets indicator at over 100% or 0% for an input.
Logging Print Prints date, time, scaling, chart speed, process variable, and engineering unit at a programmed interval. Tag Number Sets a tag number by 7 figures every channel.
List Print Prints chart speed, sensor type, measurement range, engineering unit, alarm setting value comment, printing description, logging print and on/off zone. Copy Function Copies a channel setup. 
Affix Print Prints channel number by the analog recording. Setting Input Offset etting input offset is possible for every channel. 
Dot Print Skip Skips recording of an unused channel. Zone Recording Specifies a recording area for every channel to separate into tracks.
Programming Programs chart speed, alarm setting value, logging, dot point skip, date and time. Alarm Print Prints occurrence time, occurrence channel, setting number, and alarm type in purple at occurrence of alarm.
Memory A built-in lithium battery protects the clock function backup. Alarm Recovery Print Prints recovery time, recovery channel, setting number, and alarm type in purple at recovering of an alarm. 
Alarm Sets 2 types–high and low–per channel for a total of 4 levels. Alarm Hysteresis Sets an alarm hysteresis width 0% full scale or  0.5% full scale.
Clock Indicates year, month, day, hour and minute.    


Specifications

Input Signal
Thermocouple - J, K, T, E, B, S, R, C, N, U, L, Au-Fe 
RTD - PT100, JPT100
DC Voltage - ±10mV, 0-20mV, 0-50mV, ±1V, 1-5V
Current - 4-20 mA dc, with external 250W shunt resistor

Performance
Recording Width: 100 mm calibrated
Recording Accuracy: ±0.2% ±1 digit maximum for display/printing
Input Impedance:
   mV/tc input - 10MW
   Vdc input - 1MW, mA input - 100W
Common Mode Rejection Ratio (CMRR): 140 db
Normal Mode Rejection Ratio (NMRR): 60 db
Dielectric Strength:
   Power input/ground - 1500 Vac
   Input/ground - 500 Vac
Vibration Resistance: 1 m/s2 maximum 10 - 60 Hz
Shock Resistance: 2 m/s2 maximum
Chart Feed Accuracy: ±0.1% maximum
Clock Precision: ±50 ppm

Power Source
Power Input: 85 to 264 Vac
Frequency: 45 to 65 Hz
Power Consumption: 30 VA
 

Recording and Printing
Recording: Raster-scan printing
Printing: Dotting with 6-color ribbon
Dot Print Interval: 10.0 second / 6 channel maximum
Chart Paper: Length - 52.5 ft. (16m)
Chart Speed: 28 speeds, user selectable, from 10-1500 mm/hr
Printing Colors: Purple, red, green, blue, brown, black

Alarm-Input/Output:
Outputs:
   1 relay drive per setting, up to 6 relays
   250 Vac 3A/ 30Vdc 3A/ 125Vdc 0.5A
Quantity per Channel: 4
Digital Inputs: Maximum of 3

Normal Operating Conditions
Ambient Temperature: 32° to 122°F (0° to 50°C)
Relative Humidity: 35 to 85%, non-condensing

Communications
Standard: RS-232C
Optional: RS-485 (Modbus RTU)

Structure
Dimensions: 144 × 144 × 175 mm (5.7" × 5.7" × 6.9")
Mounting: Panel mount, allowable inclination – 30°
Panel Cutout: 138 × 138 mm (5.43" × 5.43")


How To Order
The RCR-600 is offered with the options listed in the worksheet.  Create an ordering code by filling in the boxes with the appropriate number and/or letter designation for your requirements and a part number will be assigned, or choose one of the basic systems listed below.

Standard lead time is stock to 4 weeks.

Basic Systems (Part Number and Description)

Accessories
RCR40001 - 6-point dotting, 6 relay/digital outputs,
no out of paper sensor, with RS-232C data interface

RCR40002 - 6-point dotting, no relay/digital outputs,
no out of paper sensor, with RS-232C data interface

RCR40003 - 6-point dotting, 6 relay/digital outputs & 3 digital inputs, no out of paper sensor, with RS-232C data interface

RCR40004 - 6-point dotting, no relay/digital outputs,
has out of paper sensor, with RS-232C data interface
RCA40901 - Chart paper – Z fold style, 52.5 ft. (16 m)

RCA40902 - Replacement Multi-Color Ribbon

RCA40903 - Precision Shunt Resistor, 250W

RCA40904 - Portable Case